Attention! You are currently viewing old documentation:
This section of our website concerns references that are no longer manufactured or marketed.
Please use with caution.
CP2
• Turbidity limit detection • PVC probe, IP 68 • Detection range: within 100 ... 30 000 FAU • Immersion depth: Max. 10 m • To connect to a TURBISWITCH GS4
APPLICATIONS
- Level detection of blanket sludge - Phase separation detection - Effluents monitoring - Monitoring of turbidity limit of liquids
DESCRIPTION
Turbidity describes the content of particles in a liquid. Immersion probe TURBISWITCH CP2 is designed for open vessels or channels, to monitor the turbidity of the liquid. The emitter sends an infra-red light beam through the liquid to the receiver. The attenuated emitted light is measured by the control unit TURBISWITCH GS4 which switches the output relay if the pre-set value is exceeded. The detectable turbidity ranges are from about 100 up to about 30,000 FAU.
General Catalog 425-02 08/06/2022 14:10
Immersion turbidity probe TURBISWITCH CP2
BAMO Mesures SAS
22, Rue de la Voie des Bans Parc d'activités de la Gare
95100 ARGENTEUIL France