• For hot sanitary water, drinkable water • pH must be constant (for free chlorine) • Temperature: up to 50 °C or up to 70 °C • Pressure: up to 8 bar
APPLICATIONS
Concentration measurement of free chlorine or chlorine dioxide, hot sanitary water and drinkable water. Caution: The pH must be constant for free chlorine measurement.
TECHNICAL FEATURES
Measured parameter:
AS ... CL: Free chlorine
AS ... CD: Chlorine dioxide
Disinfection additives: Inorganic compounds, such as NaOCl, Ca(OCl)₂, gaseous chlorine, chlorine produced by electrolysis Chlorine dioxide: Chlorite/ Chlorine; Chlorite/ Acid Measuring system: Semi-open cell with 3 electrodes and electrolyte Output signal: 4 ... 20 mA Operating temperature:
AS2: 1 to 50°C
AS3: 1 to 70°C
Automatic temperature compensation Ambient temperature: From 0 to 55 °C Operating pressure: 8 bar max. Cleaning system Equipment: RV1 (option) Flow rate: Flow must be constant
15 l/h as a minimum, without RV1 equipment
45 l/h up to 90 l/h with RV1 equipment
pH operating range:
AS ... CL: pH must be constant
AS ... CD: between pH 1 and pH 12
Zero adjustment Not necessary (from factory) Slope calibration: Only 1 point necessary with BAMOPHAR 194:
Chlorine: DPD-1
Chlorine dioxide: DPD-1 (without chlorine)
Interferences:
AS ... CL: Ozone, chlorine dioxide and chlorite
AS ... CD: Ozone, chlorine, chlorite, measured with less than 2% of their concentration
General Catalog 193-13 03/07/2018 17:07
Free chlorine and chlorine dioxide sensors AS2 and AS3
BAMO Mesures SAS
22, Rue de la Voie des Bans Parc d'activités de la Gare
95100 ARGENTEUIL France