• Conductivity and resistivity probes • Cell factors 0.1 and 0.01 • Reliability and sturdiness • Construction: AISI 316 and PTFE, PVC • For in-line or immersion measurements
APPLICATIONS
Conductivity measurements (0.05 to 10 μS/cm) or resistivity (10 kΩ.cm to 30 MΩ.cm), on water.
- Controls of rainwater or runoff
- Controls on demineralized water in industrial process:
Boilers monitoring, etc.
Parts rinsing in galvanic process
Electronic components rinsing
DESCRIPTION
External frame surrounds the active part (centric electrode) and limit the measuring area used for measurement, ensuring perfect shield grounding.Some models have a built-in temperature sensor (Pt100 type). For a high mechanical resistance, cells are in AISI 316 L with PTFE insulation.
Cell factor
The cell factor is the ratio between the measurement done with the cell and value displayed on the monitor.
Example: A probe with a cell factor K = 0.1 immersed in a liquid of resistivity 10 kΩ.cm, measures a resistor value of 1 kΩ. The monitor will display a corrected value by factor K: resistivity displayed = 10 kΩ.cm .
For highly conductive liquids (low resistivity) it will be necessary to use probes with cell factor of K=1 or K=10 (BF1200; data sheet 361-01) or inductive sensors (toroidal sensor on data sheet 364-01/ 05).
Complete measuring system
A complete system includes: • 1 Coaxial probe with K = 0.01 or K = 0.1 • 1 Connector PL259 or direct connection to screw connectors or cable output • Coaxial cable CCA type • A 3-wire cable C3B type (when Pt 100 sensor is built-in the probe) • One BNC plug for monitor connection. • A monitor BAMOPHAR 323 (see data-sheet 323-01).
General Catalog 360-01 25/02/2022 14:03
Conductivity and resistivity probes BS - BC
BAMO Mesures SAS
22, Rue de la Voie des Bans Parc d'activités de la Gare
95100 ARGENTEUIL France