• Continuous turbidity measurement according to ISO 7027 90° scattered light measurement and 180° transmitted light measurement • Measuring cell: AISI 316 L • Measuring ranges: 0.01 ... 20 FNU 0.1 ... 500 FNU & 100 ... 1000 FAU • Output signal: 4-20 mA (4-wire) • Process connections: DIN flanges or Triclamp
APPLICATIONS
• Well water monitoring • Monitoring of membranes for filtration or reverse osmosis • Wastewater control in outlet of WWTP • Cooling plant monitoring • Filter monitoring
DESCRIPTION
TRUBOMAT GAB measures the turbidity of liquid applying nephelometric method. For this purpose, a combined transmitted & scattered beams measurements are used, in which a transmitter and a receiver face each other and another transmitter is arranged orthogonally (at an angle of 90°). The atenuated and scattered lights are recorded by the electronic module (in the connection head) and process the results to transmit a linear output signal. TRUBOMAT GAB delivers a 4-20 mA signal, according to the set measuring range.
TRUBOMAT GAB 20: 5 scales to select by DIP switches from 0.01 ... 1 up to 0.01 ... 20 FNU
TRUBOMAT GAB 1000: 5 scales to select by DIP switches from 0.1 ... 500 FNU up to 100 ... 1000 FAU
INSTALLATION Installation only in vertical piping and rising flow. Optimal condition: tranquilization distances of 600 mm before and 400 mm, after the TRUBOMAT.
General Catalog 446-01 22/10/2021 08:53
In-line turbidity system TRUBOMAT GAB
BAMO Mesures SAS
22, Rue de la Voie des Bans Parc d'activités de la Gare
95100 ARGENTEUIL France