Galvanic dissolved oxygen sensor
Sturdy construction
Easy fitting and maintenance
Range: 0...20 ppm (0-100%)
Large electrolyte volume
Easy membrane replacement
HDPE membrane
Built in temperature sensor: Pt 100 Ohm / 0°C
Construction: POM
PRINCIPLE
The sensor BAMOX 7200 uses a lead anode to improve the resolution (0.03 ppm in lower range) and a High-Density Polyethylene (HDPE) membrane; signal delivered is 45 mV at 100%.
Comparatively to a PTFE membrane (33 mV at 100%), the HDPE membrane has a quicker answer (approx. half the response time).
The BAMOX 7200 sensor is designed for wastewater applications.
TECHNICAL FEATURES
BAMOPHOX 7200 sensor
Range: 0.03...20 ppm Cathode: Silver Anode: Lead Electrolyte: special molarity, reagent grade Membrane 45 mV: HDPE Maximal temperature: 50°C Minimal fluid speed: 0.05 m/s PG colour: Grey Cable length: 3 m Temperature sensor: Pt 100 Ohm / 0°C; 2 wires
Sensor holder
Immersion probe: standard length 1 m; maximal length 3 m Construction: PVC Electrical connection: screw connectors Housing: PBT, IP65